Supercontinuum laser-based monochromatic radiant source for detector spectral responsivity characterization in the range 0.9–1.6 μm using absolute cryogenic radiometer

Autor: A. S. Baturin, A. Yu. Dunaev, S. P. Morozova, V. N. Krutikov
Rok vydání: 2020
Předmět:
Zdroj: Izmeritel`naya Tekhnika. :28-33
ISSN: 0368-1025
Popis: An improved monochromatic radiant source with spectral bandwidth of 4 nm based on supercontinuum laser and a double monochromator was included in absolute cryogenic radiometer-based facility to improve the accuracy of spectral responsivity measurement in the range 0.9–1.6 μm. The developed feedback system ensures stabilization of monochromatic radiant power with standard deviation up to 0.025 %. Radiant power that proceeds detector under test or absolute cryogenic radiometer varies from 0.1 to 1.5 mW in dependence of wavelength. The spectral power distribution of its monochromatic source for various operating mode is presented.
Databáze: OpenAIRE