A Chemical‐Stop Rapid Flow Apparatus
Autor: | John Hanacek, George H. Gibson, Richard W. Lymn |
---|---|
Rok vydání: | 1971 |
Předmět: | |
Zdroj: | Review of Scientific Instruments. 42:356-358 |
ISSN: | 1089-7623 0034-6748 |
DOI: | 10.1063/1.1685093 |
Popis: | A rapid flow apparatus with multiple mixers is described, capable of measuring reaction times of 0.008 to 0.5 sec. The solution‐containing parts are of Teflon and Lucite to minimize metal contamination. A rack and pinion drive, using an electric clutch, is described, along with a positive assay valve needed when the apparatus is used as a chemical‐stop flow apparatus. |