Radiometric calibration of the three-channel imaging polarimeter

Autor: Cheng-liang Yin, Lei Zhang, Xiao-lin Luo, Xiao-cheng Liu, Chang-pu Duan
Rok vydání: 2013
Předmět:
Zdroj: SPIE Proceedings.
ISSN: 0277-786X
DOI: 10.1117/12.2034822
Popis: Radiometric calibration is of critical importance for information quantification of imaging polarimeters. In this paper, an integral sphere which had been traced to cryogenic radiometer was used as transfer standard in our calibration facility. The linearity, uniformity, stability of our imaging polarimeter were calibrated. The combined uncertainty in the responsivity of an imaging polarimeter was about 7.5%. At last, technical proposals of reducing uncertainty budget were briefly discussed.
Databáze: OpenAIRE