Popis: |
A direct comparison of the standards for air kerma of the Bundesamt fur Eich- und Vermessungswesen (BEV), Austria, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 137Cs radiation beam of the BIPM in March 2009. The result, expressed as ratios of the BEV and the BIPM standards for air kerma, is 1.0041 with a combined standard uncertainty of 3.1 × 10−3. The result of the earlier direct comparison in 137Cs γ rays, made in 1995, was 0.9961 (28); taking into account the changes made recently to both standards, the 1995 result becomes 1.0035 (24), in agreement with the present comparison result. The degrees of equivalence between the BEV and the other participants in the key comparison have been calculated and the results are presented in the form of a matrix. A graphical presentation is also given. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA). |