Infrared backscatter imaging spectroscopy of trace analytes at standoff
Autor: | Andrew Kusterbeck, Christopher A. Kendziora, Robert Furstenberg, R. Andrew McGill, Christopher J. Breshike, Viet Nguyen |
---|---|
Rok vydání: | 2019 |
Předmět: |
010302 applied physics
Spectral signature Materials science Backscatter Infrared business.industry General Physics and Astronomy 02 engineering and technology 021001 nanoscience & nanotechnology 01 natural sciences law.invention Imaging spectroscopy Optics law 0103 physical sciences Diffuse reflection Fourier transform infrared spectroscopy 0210 nano-technology business Spectroscopy Quantum cascade laser |
Zdroj: | Journal of Applied Physics. 125:104901 |
ISSN: | 1089-7550 0021-8979 |
DOI: | 10.1063/1.5079622 |
Popis: | We are developing a proximal noncontact detection technique for trace amounts of hazardous materials using diffuse reflectance infrared spectral signatures. Spectroscopy is performed at a distance by tuning a quantum cascade laser system through its wavelength range (6-11 μm) as the active illumination source and collecting a portion of the diffusely reflected light from the target onto an MCT focal plane array. The signals from each collected frame are binned and processed into an image hypercube that contains spectral and spatial information. The primary motivation of this work is to protect the loss of life by detecting trace explosives on contaminated surfaces before attacks occur. Here, we present results from backscatter experiments on trace samples with low loading and fill factors on glass substrates and compare them with a conventional benchtop analysis technique, FTIR diffuse reflectance. The backscatter results illustrate the ability to detect explosives at 1 m distance at the nanogram level, which is beyond the capability of the benchtop FTIR diffuse reflectance measurement. |
Databáze: | OpenAIRE |
Externí odkaz: |