An improved method of Cr purification for high precision measurement of Cr isotopes by double spike MC-ICP-MS

Autor: Xiangli Wang, Lixin Zhang, Guilin Han, Guangliang Wu, Jian-Ming Zhu
Rok vydání: 2018
Předmět:
Zdroj: Journal of Analytical Atomic Spectrometry. 33:809-821
ISSN: 1364-5544
0267-9477
Popis: Chromium (Cr) isotopes have been used to trace pollution processes and reconstruct paleo-redox conditions. However, the precise determination of Cr isotopes is still challenged by difficulties in purifying Cr from samples with low Cr and high matrix content. Here, we developed an improved, flexible and easily operated three-step chromatographic procedure to separate Cr from high-matrix samples. A continuous, two-stage column (Step I) filled with 2 mL cation resin AG50W-X8 (200–400 m) and 2 mL anion resin AG1-X8 resin (100–200 m), followed by Step II, with 1 mL of AG1-X8 resin (200–400 m), was used to remove at least 99% of Ca, Fe, and Ti, and >90% of V and most other matrix elements, even for samples with Fe/Cr ≈ 6000 and Ti/Cr ≈ 1000. The recovery of Cr in both Step I and Step II can reach nearly 100%. Step III utilizes 2 mL AG1-X8 (100–200 m) combined with an (NH4)2S2O8 oxidant to remove residual matrix elements and achieve high-purity Cr. The total yield of Cr through our three-step procedure is greater than 80% even for low-Cr (2.6 μg kg−1) samples. A relatively small sample amount (300–600 ng Cr) is enough to achieve high precision Cr isotope measurement due to a low procedural blank (