Imaging dose of megavoltage computed tomography (MVCT) for treatment verification in the tomotherapy of breast cancer

Autor: Maria Y. Y. Law, K. W. Cheung, P. M. Wu, H. I. Lam, K. K. Sang, H. F. Choi, W. M. Chan, Y. W. Ho
Rok vydání: 2017
Předmět:
Zdroj: Journal of Radiotherapy in Practice. 17:244-252
ISSN: 1467-1131
1460-3969
DOI: 10.1017/s1460396917000590
Popis: AimThe purpose of this study was to investigate whether significant difference exists on radiation dose delivered to organs at risks in megavoltage computed tomography (MVCT) verification using three predefined scanning modes, namely fine (2 mm), normal (4 mm) and coarse (6 mm). This will provide information for the imaging protocol of tomotherapy for the left breast.Materials and methodsOrgan doses were measured using thermoluminescent dosimeters (TLD-100) placed within a female Rando phantom for MVCT imaging. Kruskal–Wallis test was conducted with pResultsStatistically significant difference existed in organ absorbed dose between different scan mode selections (pConclusionsOptimisation of imaging protocols is of paramount importance to keep the radiation exposure ‘as low as reasonably achievable’. The recommendation of undergoing daily coarse mode for MVCT verification in breast tomotherapy not only mitigates the radiation exposure to normal tissues, but also trims the scan-acquisition time.