Optimization of isolated electron pulses in the picosecond range from a linear accelerator using a streak camera–TV diagnostic system
Autor: | G. R. Tripp, Klaus H. Schmidt, G. S. Mavrogenes, C. Jonah, L. W. Coleman, S. Gordon |
---|---|
Rok vydání: | 1976 |
Předmět: | |
Zdroj: | Review of Scientific Instruments. 47:187-189 |
ISSN: | 1089-7623 0034-6748 |
DOI: | 10.1063/1.1134587 |
Popis: | We describe the technique and the results of a measurement of the width of a fine structure pulse from our L‐band electron linear accelerator. The pulse width was measured by measuring the width of the Cerenkov light pulse generated by the electron beam using a streak camera–TV camera–computer system. With very low beam currents, a pulse width of 20 psec (FWHM) was measured, while at higher beam currents a pulse width of 27 psec was measured showing a space‐charge effect. In addition, a broad energy spectrum for the electrons or improper focusing into a 270° bending magnet was also shown to increase the pulse width of the beam after it had traversed the magnet. |
Databáze: | OpenAIRE |