The effect of roller head pump on platelet deterioration during the simulated extracorporeal circulation

Autor: Yukinori Kozuma, Tatsuya Furugaki, Bryan J. Mathis, Hiroaki Sakamoto, Toru Tsukada, Tomomi Nakajima, Yasuyuki Suzuki, Yuji Hiramatsu, Osamu Shigeta
Rok vydání: 2020
Předmět:
Zdroj: Journal of Artificial Organs. 24:22-26
ISSN: 1619-0904
1434-7229
DOI: 10.1007/s10047-020-01192-x
Popis: Roller pumping results in hemolysis and adverse effects on coagulation, but there are few reports on the influence of roller heads on platelets. Here, we evaluate the interaction between roller pumping and platelet function using a simulated extracorporeal circuit incorporating a vinyl chloride tube and roller head pump with 30 min recirculation. Platelet aggregation, platelet count, microparticle, P-selectin, Phosphatidylserine (PS) exposure and Ricinus Communis Agglutinin 1 (RCA-1) were measured before, 5, 10, 20, and 30 min after the recirculation using 100 ml of fresh human blood that had obtained from healthy volunteers (n = 9). Platelet aggregation and platelet count gradually decreased but microparticles significantly increased after the recirculation (P