Geopotential measurement with a robust, transportable Ca+ optical clock

Autor: Mengyan Zeng, Yanmei Hao, Yuzhuo Wang, Kelin Gao, Baolin Zhang, Huaqing Zhang, Yige Lin, Shiying Cao, Qunfeng Chen, Fang Zhanjun, Yao Huang, Tianchu Li, Kun Liang, Hua Guan, Fang Fang
Rok vydání: 2020
Předmět:
Zdroj: Physical Review A. 102
ISSN: 2469-9934
2469-9926
DOI: 10.1103/physreva.102.050802
Popis: We present a robust, transportable ${\mathrm{Ca}}^{+}$ optical clock, with a systematic uncertainty of $1.3\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$ limited by the black-body radiation (BBR) field evaluation and an uptime rate of $g75%$ over a 20-day period. The clock is then installed in an air-conditioned car trailer, making it more convenient for applications. Referenced to a stationary laboratory clock, geopotential measurements are made with the transportable clock with a total uncertainty of 0.33 m (statistically 0.25 m and systematically 0.22 m) and agree with the spirit level measurement. After being moved $g1200$ km, the absolute frequency of the ${\mathrm{Ca}}^{+}$ optical clock transition is measured as 411 042 129 776 400.41(23) Hz, with a fractional uncertainty of $5.6\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}16}$, which is about one order of magnitude smaller than our previous measurement. The transportable built can be used for sub-meter-level elevation measurements, comparing intercontinental optical clocks, verifying basic physical theories, etc.