Effort on calibration of infrared spark ablation of copper with synthetic copper standards
Autor: | Stefan Seeger, Julia Wienold, Heinrich Kipphardt, Sebastian Recknagel, Harald Bresch, Heike Traub |
---|---|
Rok vydání: | 2011 |
Předmět: |
Laser ablation
Materials science Infrared medicine.medical_treatment Analytical chemistry chemistry.chemical_element Ablation Copper Atomic and Molecular Physics and Optics Analytical Chemistry chemistry Particle-size distribution medicine Calibration Particle size Instrumentation Inductively coupled plasma mass spectrometry Spectroscopy |
Zdroj: | Spectrochimica Acta Part B: Atomic Spectroscopy. 66:432-438 |
ISSN: | 0584-8547 |
DOI: | 10.1016/j.sab.2011.04.005 |
Popis: | Two types of copper samples, compact certified copper reference materials and calibration samples prepared from liquid doped, pressed copper powders, were studied in terms of accuracy of obtained calibration functions originating from infrared spark ablation. Additionally, corresponding particle size distributions of the aerosols from infrared spark ablation were recorded. It is shown that the differences in quantification results, originating from the two sets of calibration functions, could not mainly be ascribed to different particle size distributions of the two copper sample types. Possible other causes, as different ablation rates, parts of melting and differences of the chemical constitutions of the two sample types were explored. |