Simultaneous Multiwavelength Detection System for Liquid Chromatography
Autor: | Leon N. Klatt |
---|---|
Rok vydání: | 1979 |
Předmět: | |
Zdroj: | Journal of Chromatographic Science. 17:225-235 |
ISSN: | 1945-239X 0021-9665 |
DOI: | 10.1093/chromsci/17.4.225 |
Popis: | A simultaneous multi-wavelength detection system for liquid chromatography is described. The system consists of a rapid scan sectrometer with a uv silicon vidicon target as radiant energy detector, a minicomputer, a programming language, and the requisite chromatographic hardware. Algorithms required for a real-time data processing capability are described. Multi-wavelength data retained during a chromatographic separation are summarized as an absorbance-wavelength-time surface. The ability to identify components unresolved by the chromatographic column is demonstrated. Solutes unresolved by the column can be quantitated with a precision and accuracy equal to that observed for solutes resolved by the column. The use of this detection system to qualitatively assess compositional changes in a coal derived oil as a function of process, as well as the limitations of the system and required improvements are discussed. 11 figures, 4 tables. |
Databáze: | OpenAIRE |
Externí odkaz: |