Advantage of fan beam collimators for contrast recovery of hyperfixation in clinical SPECT

Autor: O. de Dreuille, J.L. Stievenart, Bernard Bendriem, André Syrota, M. H. Bourguignon, B. Bok, J.F. Menard, C. Parc, Isabelle Gardin, Pierre Vera
Rok vydání: 1997
Předmět:
Zdroj: IEEE Transactions on Nuclear Science. 44:83-89
ISSN: 1558-1578
0018-9499
DOI: 10.1109/23.554830
Popis: The influence of the collimator on the contrast recovery of hyperfixation was studied using a dual-headed single photon emission computed tomography (SPECT) system with standard clinical acquisition parameters. Three parallel collimator sets and two fan beam collimator sets were tested with a Jaszczak phantom. The six spheres of the phantom were filled with /sup 99m/Tc, and four background levels were progressively obtained by adding radioactivity to the cylinder of the phantom, providing four hyperfixation levels. The effects of angular sampling and reconstruction filters have been tested. The statistical analysis was performed with analysis of variance (ANOVA). This study demonstrates the advantage of ultra-high resolution fan beam collimators for contrast recovery of hyperfixation with SPECT when using 64 projections over 360/spl deg/, in particular when the contrast is low. The authors also demonstrate that fan beam collimators permit smaller size hyperfixation detection.