High dose efficiency, ultra-high resolution amorphous selenium/CMOS hybrid digital X-ray imager

Autor: Peter M. Levine, Christopher C. Scott, Abdallah El-Falou, Alireza Parsafar, Karim S. Karim
Rok vydání: 2015
Předmět:
Zdroj: 2015 IEEE International Electron Devices Meeting (IEDM).
DOI: 10.1109/iedm.2015.7409803
Popis: We demonstrate measured X-ray dose efficiency results from a 5.6-μm × 6.25-μm pixel-pitch direct-conversion amorphous selenium/CMOS hybrid X-ray imager. Compared to existing scintillator-based imagers, our approach enables up to 100x gains in DQE at spatial frequencies of 20–60 cycles/mm, which could radically accelerate bioengineering research. The measured MTF is 50% at 32 cycles/mm, corresponding to a 16-μm effective object size. We also demonstrate a MTF model which confirms experimental results with an RMS error of 0.02.