A fast recovery pulsed nuclear magnetic resonance sample probe using a delay line
Autor: | M. Engelsberg, I. J. Lowe |
---|---|
Rok vydání: | 1974 |
Předmět: | |
Zdroj: | Review of Scientific Instruments. 45:631-639 |
ISSN: | 1089-7623 0034-6748 |
DOI: | 10.1063/1.1686702 |
Popis: | A sample probe for pulsed nuclear magnetic resonance that replaces the normal resonant circuit by a lumped parameter delay line is described. Theoretical analysis shows that the delay line probe has a signal‐to‐noise ratio and conversion efficiency of power into rotating magnetic field equivalent to a resonant circuit of Q =(2π) (delay time) (Larmor frequency). However, the delay line probe has a much wider bandwidth, much shorter transient decay time, and faster recovery from the rf excitation pulse. Experimental results are included that demonstrate submicrosecond recovery times for the observation of the signal. |