Autor: |
B. Becker, S. Medved, Y. Diawara, C. Hervé, Dina Carbone, Peter Boesecke, T. Thorson, Roger D. Durst, G. Wachter, Bob B. He, V. Sedov, D. Khazins, Menyhert Kocsis |
Rok vydání: |
2006 |
Předmět: |
|
Zdroj: |
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 563:172-176 |
ISSN: |
0168-9002 |
DOI: |
10.1016/j.nima.2006.01.117 |
Popis: |
We describe in this paper the performance under synchrotron radiation of a sealed gaseous detector, Vantec-1, operating at high pressure and less prone to discharges. It is consequently operating at high gain (>105) at high local counting rates up to 106 cps/mm2. To achieve these characteristics, a thin resistive layer is used on a glass plate between the mesh and the readout anode. The performance achieved with this detector is suited for applications in X-ray diffraction and synchrotron radiation. This detector has shown high reliability over time under harsh synchrotron radiation environment and robustness in manufacturing environment. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|