Autor: |
J. Cronin, M. Grattan, N. Agazaryan, Dan Sporea, Laura Mihai, P. Woulfe, Elfed Lewis, Anand P. Santhanam, Sinead O'Keeffe, D. McCarthy, Alan R. Hounsell |
Rok vydání: |
2014 |
Předmět: |
|
Zdroj: |
SPIE Proceedings. |
ISSN: |
0277-786X |
DOI: |
10.1117/12.2058572 |
Popis: |
A PMMA based plastic optical fibre sensor for use in real time radiotherapy dosimetry is presented. The optical fibre tip is coated with a scintillation material, terbium-doped gadolinium oxysulfide (Gd2O2S:Tb), which fluoresces when exposed to ionising radiation (X-Ray). The emitted visible light signal penetrates the sensor optical fibre and propagates along the transmitting fibre at the end of which it is remotely monitored using a fluorescence spectrometer. The results demonstrate good repeatability, with a maximum percentage error of 0.5% and the response is independent of dose rate. |