Traceable point-like 68Ge/68Ga source with a spherically symmetric positron absorber for PET scanners

Autor: Takahiro Yamada, Hiroki Miyatake, Hidetake Ishizu, Mio Okamoto, Tomoyuki Hasegawa, Kei Kikuchi, Takahiro Mikamoto, Yasushi Sato, Yusuke Inoue
Rok vydání: 2020
Předmět:
Zdroj: Radiological Physics and Technology. 13:170-176
ISSN: 1865-0341
1865-0333
Popis: A novel traceable point-like 68Ge/68Ga source was developed for calibration of positron emission tomography (PET) scanners. The source was equipped with a spherically symmetric acrylic positron absorber. The physical characteristics of photons emitted from the point-like source were evaluated by Monte Carlo simulation, considering possible geometrical uncertainties. The calibration factor of a clinical PET/CT scanner was evaluated using four manufactured point-like sources as a practical application of the point-like source. The results were consistent with that determined by the conventional cross-calibration method. The traceable point-like 68Ge/68Ga source is expected to be a simple and practical tool for determining the calibration factor and evaluating the physical characteristics of PET scanners.
Databáze: OpenAIRE