Tissue-equivalent radiochromic waveguide dosimeters for X-and gamma rays and fast neutrons

Autor: William L. McLaughlin, Stanley Kronenberg, Carl R. Siebentritt
Rok vydání: 1986
Předmět:
Zdroj: International Journal of Radiation Applications and Instrumentation. Part D. Nuclear Tracks and Radiation Measurements. 12:561-564
ISSN: 1359-0189
DOI: 10.1016/1359-0189(86)90651-5
Popis: Radiochromic waveguide dosimeters consist of small fluorinated ethylene propylene (FEP) tubing containing a radiation-sensitive organic liquid or gel, which, because of its higher refractive index, propagates light through the tubing as an optical waveguide. The measurement of dose is accomplished by reading radiation induced changes in the ratio of transmittances at two visible light wavelengths. These dosimeters read the gamma and neutron tissue doses. Real time dosimeters were built on this principle for use in radiation therapy.