A Mixture of Glycosylated proBNP and BNP-32 Is a Suitable Calibrator for BNP Immunoassays
Autor: | Norio Ota, Hiroshi Nishi, Toshio Nishikimi, Masayo Igarashi, Kazukiyo Horii, Keiichi Masuta, Yasuaki Nakagawa, Kenzo Funatsuki, Izuru Masuda, Youko Inoue |
---|---|
Rok vydání: | 2019 |
Předmět: |
Immunoradiometric assay
Chromatography Plasma samples business.industry medicine.drug_class cardiovascular system Automated analyzer Natriuretic peptide Medicine Potential source cardiovascular diseases business human activities Blood stream hormones hormone substitutes and hormone antagonists circulatory and respiratory physiology |
Zdroj: | Open Journal of Clinical Diagnostics. :51-63 |
ISSN: | 2162-5824 2162-5816 |
DOI: | 10.4236/ojcd.2019.92004 |
Popis: | Aim: Measurement of B-type natriuretic peptide (BNP) is widely used as a diagnostic and risk assessment tool for cardiovascular disease. Recent studies have demonstrated that BNP-32 and its precursor proBNP circulate in the blood stream, and that most commercial BNP immunoassays measure both forms. However, recombinant or synthetic BNP-32 is used as the standard for those BNP immunoassays. This gap between clinical samples and the standard might be a potential source of variation in BNP measurements among assays. The purpose of this study is to validate a more suitable calibrator for BNP immunoassays. Methods: External BNP calibrators containing both BNP-32 and glycosylated proBNP were prepared at two concentration levels. Target BNP concentrations of the low and high levels were 40 and 160 pg/mL, respectively. And to reflect clinical samples, the molar ratios of BNP-32 to glycosylated proBNP in these concentration levels were adjusted to 50:50 and 25:75, respectively. BNP concentrations of plasma samples along with the external BNP calibrators were measured at two commercial labs and using an automated analyzer MI02 Shionogi® BNP (MI02). These samples and the calibrators were also measured using an immunoradiometric assay (IRMA) as a standard assay procedure. Concentrations of the plasma samples measured at the labs or using the MI02 were adjusted according to a comparison of the measured concentrations of the external BNP calibrators with the IRMA. Results: After measured concentrations of the plasma samples were adjusted using the external BNP calibrators, the correlation between each measurement and the IRMA was improved. The range of the slopes according to Passing-Bablok regression analysis narrowed from 0.628 - 0.955 to 0.911 - 1.005. Conclusions: Our data suggests that a mixture of BNP-32 and glycosylated proBNP at different ratios by concentration level is suitable for calibration to minimize variations in BNP measurements among immunoassays. |
Databáze: | OpenAIRE |
Externí odkaz: |