Proficiency Investigation of Torque Tools Calibration as a National Demand

Autor: K.M. Khaled, Seif M. Osman
Rok vydání: 2019
Předmět:
Zdroj: MAPAN. 34:207-215
ISSN: 0974-9853
0970-3950
DOI: 10.1007/s12647-019-00301-3
Popis: The current paper represents a round loop for calibration of torque tools organized by NIS. These measurements are aimed to evaluate the performance of torque tools calibration laboratories in Egypt in the range from 200 N m up to 1000 N m in both clockwise and anticlockwise directions. Results are evaluated using En-value method and Z-score method based on the weighted means and the corresponding uncertainties. The outlier test and consistency test were performed on the measurements before applying the Z-score test. The Z-score evaluation shows that participants’ results are consistent specifically in the range from 400 N m up to 1000 N m. Further comparison with lower capacity artifact is recommended in the lower torque range.