Proficiency Investigation of Torque Tools Calibration as a National Demand
Autor: | K.M. Khaled, Seif M. Osman |
---|---|
Rok vydání: | 2019 |
Předmět: | |
Zdroj: | MAPAN. 34:207-215 |
ISSN: | 0974-9853 0970-3950 |
DOI: | 10.1007/s12647-019-00301-3 |
Popis: | The current paper represents a round loop for calibration of torque tools organized by NIS. These measurements are aimed to evaluate the performance of torque tools calibration laboratories in Egypt in the range from 200 N m up to 1000 N m in both clockwise and anticlockwise directions. Results are evaluated using En-value method and Z-score method based on the weighted means and the corresponding uncertainties. The outlier test and consistency test were performed on the measurements before applying the Z-score test. The Z-score evaluation shows that participants’ results are consistent specifically in the range from 400 N m up to 1000 N m. Further comparison with lower capacity artifact is recommended in the lower torque range. |
Databáze: | OpenAIRE |
Externí odkaz: |