Niezwykle wysoka amplituda sygnału pęczka Hisa w kanale przedsionkowym ICD
Autor: | Magdalena Sokołowska, Aleksandra Gajek, Agnieszka Sławuta, Martek Frączkowski, Jacek Gajek, Jakub Adamowicz |
---|---|
Rok vydání: | 2017 |
Předmět: | |
Zdroj: | Folia Cardiologica. 12:295-297 |
ISSN: | 2353-7760 2353-7752 |
DOI: | 10.5603/fc.2017.0057 |
Popis: | In patients with chronic atrial fi brillation and congestive heart failure presenting with narrow QRS complex the direct His-bundle pacing the use of dual chamber implantable cardioverter-defibrillator can be a defi nite treatment option meeting all therapeutic goals. This approach enables heart rate regularization, appropriate increase of beta-blocker dose with additional benefi t in terms of arrhythmic death prophylaxis. The proper functioning of implantable cardioverter- -defibrillator demands appropriate atrial channel/His-bundle signal sensing. The low His-bundle electrogram amplitude usually does not interfere with the device function and the DDI or DVI programming modes are reasonable. The authors present a case of unusually high His-bundle signal amplitude sensed by atrial channel of implantable cardioverter-defibrillator, which could potentially contribute to some particular programming considerations. The attending physician should be aware of some arrhythmia discriminators which may delay or even withhold the appropriate shock. |