Investigation of Energy Responses of Germanium Detectors and Correction of Measured Spectra by Means of Monte Carlo Simulation

Autor: Kunio Doi, Thomas R. Fewell, Chin-Tu Chen, Heang Ping Chan, Ralph E. Shuping
Rok vydání: 1984
Předmět:
Zdroj: Radiation Research. 99:443
ISSN: 0033-7587
DOI: 10.2307/3576321
Popis: A Monte Carlo program was developed which determines the energy response of X-ray detectors of various "nested" sizes simultaneously. Twenty-one planar germanium detectors with cylindrical crystals were chosen, which are commercially available for X-ray spectrometry. The photopeak efficiency, K alpha and K beta escape fractions. Compton fraction, elastic fraction, penetration fraction, and the Compton continuum were tabulated as a function of incident X-ray energies from 12 to 300 keV in 2-keV increments. Coefficients of polynomials fitted to the Compton continua were also calculated to facilitate computer implementation of spectrum corrections. Spectrum correction methods which use the Compton continua determined by Monte Carlo calculation were compared with those approximated by rectangles. Comparison of the residual root-mean-square (RRMS) values between the incident and corrected spectra showed that the former method yielded lower RRMS values. However, the differences in these values are small, indicating that the simple rectangular approximation may be adequate for most spectral measurements.
Databáze: OpenAIRE