Primary Total Hip Arthroplasty With Duraloc Cup in Patients Younger Than 50 Years

Autor: Tian Jialiang, Shen Bin, Yang Jing, Mu Zhongyou, Pei Fuxing, Zhou Zongke
Rok vydání: 2009
Předmět:
Zdroj: The Journal of Arthroplasty. 24:1184-1187
ISSN: 0883-5403
DOI: 10.1016/j.arth.2008.12.002
Popis: We evaluated the clinical and radiographic results of 67 (64 patients) cementless Duraloc-300 cups for young active patients after a mean follow-up period of 6.2 years. The preoperative mean Harris hip score improved from 46.24 to 96.5 points at 5 years. The survivorship of the cup, using radiographically confirmed aseptic loosening as the end point, was 100% at 5 years. The mean rate of liner wear was 0.125 mm/y (00-0.39 mm/y). Acetabular osteolysis was found in 14% (9 hips) of the 67 hips, and the osteolysis is related to polyethylene wear (P = .0024) and sex (P = .001). Although there was no aseptic loosening of the components, there was a high rate of liner wear of the polyethylene liner and acetabular osteolysis.