Mobile versus fixed automated external defibrillators (AED deployment in a geographically dispersed population: analysis of the girona territori cardioprotegit project

Autor: M Salvans Cirera, P Loma-Osorio Ricon, M Pons Monne, P Llongueras Espi, D Vinas Fernandez, R Brugada Tarradellas, J Aboal Vinas, Mahipal Singh, H Kassem Perez, E Badosa Gali, J Pascual Mayans, M. Nunez Torras, R Ramos Polo, F Graterol Torres
Rok vydání: 2021
Předmět:
Zdroj: European Heart Journal. 42
ISSN: 1522-9645
0195-668X
DOI: 10.1093/eurheartj/ehab724.1544
Popis: Introduction/Aim Public defibrillation doubles out-of-hospital cardiac arrest survival. However, the best way to provide public defibrillation coverage to geographically dispersed populations remains unknown. The aim of this study is to compare usage rates and effectivity between mobile versus fixed Automated External Defibrillators (AED). Methods This project is a prospective registry of the usage rate of public AED (542 fixed AED, 241 mobile AED) and the analysis of the electrocardiographic traces, from June 2011 until December 2019. We compared the usage rate, the proportion of shockable rhythms and defibrillation success between fixed versus mobile AED. Results Of 566 registered usages, we obtained 494 electrocardiographic traces, of which 108 (21%) were from fixed AED. The usage rate of fixed and mobile AED were 0.022use/AED-year and 0.177use/AED-year respectively. In Fixed AED group we observed a higher proportion of shockable rhythms (34.2% vs. 20.3%, p=0,01) and higher defibrillation success (79% vs. 63%, p=0,02). The proportion of patients with shockable rhythms who were transferred to a hospital were 62.1% and 50% in Fixed AED and Mobile AED group respectively (p=0,306). Conclusions In Fixed AED group we observed more shockable rhythms and higher defibrillation success rates. Mobile AED were 8 times more used. Funding Acknowledgement Type of funding sources: None.
Databáze: OpenAIRE