Autor: |
Bronislaw K. Dichter, Gary E. Galica, Frederick A. Hanser, Douglas B. Carssow, Chadwick D. Lindstrom, James D. Sullivan |
Rok vydání: |
2011 |
Předmět: |
|
Zdroj: |
SPIE Proceedings. |
ISSN: |
0277-786X |
DOI: |
10.1117/12.893814 |
Popis: |
The Teledyne microdosimeter is a novel miniature dosimeter that has become recently available to satellite manufacturers and programs to provide awareness of the total radiation dose received by the satellite and its associated subsystems. A characterization of the response of the dosimeter to protons of energies from 30 - 200 MeV as a function of angle, energy and dose rate is presented in this paper. In addition, the response of the dosimeter to a simulated Solar proton event with several different levels of shielding has been measured. These results show that the dosimeter response is relatively uniform over a wide range of conditions for protons. Monte Carlo modeling of the dosimeter for isotropic particle fluxes (both electrons and protons) has also been accomplished. It is shown that a simplified model is appropriate in determining the response of the dosimeter when using it to design low cost, simple instruments for space weather and situational awareness applications. |