In vivo fracture of a reverse total shoulder replacement humeral tray: A case report
Autor: | Trevor C. Gascoyne, Peter B. MacDonald, Eric Bohm, Martin J. Petrak |
---|---|
Rok vydání: | 2016 |
Předmět: | |
Zdroj: | Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 230:1141-1147 |
ISSN: | 2041-3033 0954-4119 |
DOI: | 10.1177/0954411916676217 |
Popis: | Failure analysis was performed for two humeral tray components which fractured in vivo after only 6 and 9 months. Implant retrieval analysis indicated that the components failed due to fatigue failure initiating from a small radius fillet at the interface of the trunnion and tray regions. Finite element simulations revealed the small radius fillet to have resulted in a large stress concentration and confirmed the possibility for fatigue failure in 6 months. The stress concentration caused by both the small radius fillet and the insufficient tray thickness contributed to the premature fatigue failure of the humeral trays. |