Problems in Analysis of Wide Boiling Range Materials Using A Gas Liquid Chromatography-Flame Ionization Detector System

Autor: H. B. Jensen, R. E. Poulson
Rok vydání: 1971
Předmět:
Zdroj: Journal of Chromatographic Science. 9:300-306
ISSN: 1945-239X
0021-9665
DOI: 10.1093/chromsci/9.5.300
Popis: The Bureau of Mines investigated the sources of error of a GLC-FID system in quantitative analysis of wide-boiling-range mixtures such as are encountered in simulated distillation of crude shale oil or petroleum. The sample-introduction system, column effects, and the FID itself were studied using normal alkanes as a diagnostic aid. The on-column inlet technique of sample introduction is compared with the flash-vaporization inlet technique. Column-adsorption effects and the use of tail-reducing materials are discussed. Problems with the FID relating to insulator and electrode contamination, electrode emissivity, sample capacity, and burner design are illustrated and discussed. Normal alkane FID response relative to n-heptane over the C/sub 7/--C/sub 32/ range is shown to be 1.00 +- 0.03 for a parallel-plate FID when operating normally.