Anodic polarization measurements of orthopaedic implant alloys in bovine serum albumin
Autor: | Takao Yamamuro, Yoshihiko Kotoura, Yuichiro Nakayama, Kouichiro Shimizu, Junji Kawafuku, P. Kumar, Takahiro Takashima, Masanori Oka |
---|---|
Rok vydání: | 1990 |
Předmět: |
Materials science
biology medicine.diagnostic_test Metallurgy Alloy technology industry and agriculture General Engineering chemistry.chemical_element General Medicine engineering.material equipment and supplies Nickel chemistry Nickel titanium Spectrophotometry biology.protein engineering medicine Bovine serum albumin Inductively coupled plasma Polarization (electrochemistry) Nuclear chemistry Titanium |
Zdroj: | Journal of Applied Biomaterials. 1:307-313 |
ISSN: | 1549-9316 1045-4861 |
DOI: | 10.1002/jab.770010406 |
Popis: | Anodic polarization measurements of several orthopaedic implant alloys—AISI 316L stainless steel, CoCrMo alloy (ASTM F-75), nickel-titanium (NiTi) alloy, commercial pure titanium, and Ti6Al4V—were carried out in Ringer's solution containing 1% bovine serum albumin (BSA) and in simple Ringer's solution. Constituent metals in the medium were analyzed by inductively coupled plasma emission spectrophotometry after measurements. Pitting potentials and total anodic current of former two alloys were constant in the two mediums and not affected by the addition of bovine serum albumin. However, the amount of dissolved metals from pure nickel and AISI 316L stainless steel were increased by the addition of BSA; dissolved metals from CoCrMo alloy were decreased by it. |