Continuous in Situ Cyanide Monitoring Using a Highly Sensitive and Selective FIA System

Autor: Yoshiko Arikawa, Takuo Akimoto, Kenji Hayashi, Hong-Seok Ji, Kaori Nagakubo, Yoko Nomura, Scott McNiven, Isao Karube, Atsushi Watanabe
Rok vydání: 2000
Předmět:
Zdroj: Environmental Science & Technology. 34:2618-2622
ISSN: 1520-5851
0013-936X
DOI: 10.1021/es991039a
Popis: A flow injection analysis (FIA) system incorporating a gas-diffusion membrane was fabricated for the detection of cyanide anion in aqueous samples. The principle of measurement is based on the reaction of o-phthalaldehyde (OPA) and cyanide in the presence of glycine to produce a fluorescent isoindole derivative. The cyanide concentration of the samples is thus proportional to the observed fluorescence intensity. Although extremely low levels of cyanide could be determined using this system (lower detection limit 0.4 ng mL-1 of CN-), measurements were affected by the presence of sulfite ion and thiols. Therefore, a gas-diffusion membrane was incorporated into the system to separate gaseous hydrogen cyanide from interferents in the sample. Consequently, this system displayed high selectivity for cyanide. The presence of sulfite ion (1 μg mL-1), 2-mercaptoethanol (0.1 μg mL-1), or 2-mercaptopropionic acid (10 μg mL-1) did not significantly increase the fluorescence intensity of cyanide solutions (16 ng mL-1)...