Continuous-flow enzyme immunoassay for thyroxine in serum
Autor: | N J Tarsa, G DiBenedetto, J P Nolan |
---|---|
Rok vydání: | 1981 |
Předmět: | |
Zdroj: | Clinical Chemistry. 27:738-741 |
ISSN: | 1530-8561 0009-9147 |
DOI: | 10.1093/clinchem/27.5.738 |
Popis: | We have evaluated a fully automated thyroxine assay involving the use of a homogeneous enzyme assay and a Technicon AutoAnalyzer II continuous flow system. Comparison by regression analysis with a thyroxine radioimmunoassay kit method gave a slope of 0.82 and a y-intercept of 7.81 micrograms/L (SE 0.86 microgram/L, r = 0.95). Within-run precision yielded CVs of 1.0-6.1%, between-day CVs were 2.0-14.4%; within-day precision showed a mean variance of 7.81 micrograms/L. Mean analytical recovery was 96.1%. Bilirubin, hemoglobin, and lipemia interfered with quantitation of results when their concentrations exceeded 50 mg/L, 300 mg/L, and 5 g/L, respectively. The reference interval for euthyroid status was calculated to be 50-110 micrograms/L. Sensitivity was 5.0 micrograms/L with a mean carryover of 1.65%. Current reagent and labor costs for enzyme immunoassay ($0.40) were less than for the manual radioimmunoassay procedure ($0.40) were less than for the manual radioimmunoassay procedure ($1.70). The assay is economical, simple to perform, and amenable to high-throughput thyroid screening in the routine laboratory. |