Response of LYSO:Ce scintillation crystals to low energy gamma-rays
Autor: | Mikhail Korzhik, Z. Usubov, V. V. Shevtsov, A.N. Shalyugin, I. F. Emeliantchik, V. Yu. Baranov, J. Budagov, V. Glagolev, K. G. Afanaciev, D. Yu. Kozlov, V. V. Tereschenko, A. A. Fedorov, V. Mechinsky, A. V. Simonenko, A. Artikov, M. A. Batouritski, R. M. Djilkibaev, Yu.I. Davydov |
---|---|
Rok vydání: | 2015 |
Předmět: |
Physics
Nuclear and High Energy Physics Range (particle radiation) Radiation business.industry Gamma ray Analytical chemistry Linearity Atomic and Molecular Physics and Optics Particle detector Lyso Crystal Optics Scintillation counter Measuring instrument Radiology Nuclear Medicine and imaging business |
Zdroj: | Physics of Particles and Nuclei Letters. 12:319-324 |
ISSN: | 1531-8567 1547-4771 |
DOI: | 10.1134/s1547477115020028 |
Popis: | The LYSO crystals of dimensions 30 × 30 × 130 mm and 10 × 10 × 10 mm are tested with different gamma sources. Energy resolutions and linearity of the energy responses in the energy range 511–1333 keV are measured. It is found that crystals have good linearity of the energy response in the entire tested energy range. Longitudinal light response non-uniformity of the long crystal is found to be less than 3.2% within the 30–120 mm range along the crystal axis. |