A clinical evaluation of an oscillometric blood pressure monitor on anesthetized horses

Autor: K.R. Branson
Rok vydání: 1997
Předmět:
Zdroj: Journal of Equine Veterinary Science. 17:537-540
ISSN: 0737-0806
DOI: 10.1016/s0737-0806(97)80226-2
Popis: The accuracy of an oscillometric blood pressure monitor (Vet/BP 6000, Sensor Devices, Inc., Waukesha, WI) was assessed in 13 anesthetized horses undergoing a variety of clinical procedures. The oscillometric blood pressure measurements using a cuff placed around the base of the tail were compared to simultaneously recorded direct pressure readings. One hundred nine sets of blood pressure measurements were recorded. In addition the pulse rates, as determined by digital palpation and oscillometric measurement, the time required for an oscillometric measurement, and the number of attempts required by the oscillometric unit were recorded. The average difference between the direct and oscillometric blood pressures were 18, 9, and 11 mmHg for the systolic, diastolic and mean arterial blood pressures, respectively. The oscillometric and direct pressure measurements showed significant correlation with a correlation coefficient of 0.93 for the mean arterial pressure. The oscillometrically determined pulse rate (average-33 pulses/min) was significantly different than the pulse rate determined by palpation (average-33.4 pulses/minute) but the difference was less than one. The average time required for a determination was 69 seconds and the average number of attempts required was 1.02 per determination.