Scintillation neutron detectors based on solid-state photomultipliers and lightguides
Autor: | A. A. Stolyarov, S. K. Karaevsky, Ravil A. Sadykov, D. N. Trunov, V. N. Marin, V. S. Litvin, S. N. Axenov |
---|---|
Rok vydání: | 2016 |
Předmět: |
010302 applied physics
Physics Scintillation Photomultiplier Physics::Instrumentation and Detectors 010308 nuclear & particles physics business.industry Astrophysics::High Energy Astrophysical Phenomena Nuclear Theory Detector General Chemistry Neutron radiation Condensed Matter Physics 01 natural sciences Wavelength Optics Silicon photomultiplier 0103 physical sciences Scintillation counter Neutron detection High Energy Physics::Experiment General Materials Science Nuclear Experiment business |
Zdroj: | Crystallography Reports. 61:106-110 |
ISSN: | 1562-689X 1063-7745 |
DOI: | 10.1134/s1063774516010090 |
Popis: | Neutron detectors based on scintillation screens ZnS(Ag)/LiF and solid-state photomultipliers have been developed. Lightguides are used to collect light. The application of a coincidence scheme provides a low dark count and a neutron detection efficiency as high as 70%. A scheme of x-y neutron detector based on wavelength shifting fibers is also proposed. Tests of the proposed versions of detectors in a neutron beam have shown their efficiency. |
Databáze: | OpenAIRE |
Externí odkaz: |