Brachytherapy source calibration, reviews, and consistency of192Ir high-dose rate afterloading sources supplied over the period of 10 years: A retrospective analysis

Autor: Vivek Kaushal, Balasubramanian Nagappan, Yogesh Kumar, Anil Kumar Dhull, Narayan P Patel
Rok vydání: 2015
Předmět:
Zdroj: Radiation Protection and Environment. 38:144
ISSN: 0972-0464
DOI: 10.4103/0972-0464.176158
Popis: Measurement and verification of strength of monomodal high-dose rate (mHDR)192 Ir source supplied by the vendor is a major part of quality assurance program. Reference air kerma rate (RAKR) or air kerma strength (AKS) is the recommended quantity to specify the strength of gamma emitting brachytherapy sources. Physicist in our institution performed the source calibration as soon as each 192 Ir new source was loaded on the mHDR afterloading machine. The AKS accurately measured using a physikalisch technische werkstatten (PTW) re-entrant chamber-electrometer system in a scatter-free geometry was used to compute the air kerma rate (AKR) at one-meter distance in the air. To ensure accurate dose delivery to brachytherapy patients, measured AKS or RAKR should be entered correctly in both HDR treatment console station (TCS) as well as treatment planning system (TPS) associated with it. The clinical outcome mainly depends not only on the accuracy of the source strength measurement in the hospital but also on the correct source strength entered into both TCS and TPS software. A retrospective study on 22 mHDR V2 sources supplied by the vendor for the period of 10 years was taken up to access the accuracy of source strength supplied to the Radiotherapy department. The results are analyzed and reported. The accuracy in measured RAKR of all 22 sources supplied by vendor was well within the tolerance limits set by the national regulatory body and international recommendations. The deviations observed between measured RAKR versus manufacturer's quoted RAKR were in the range from −1.71% to +1.15%. In conclusion, the measured RAKR have good agreement with vendor quoted RAKR values.
Databáze: OpenAIRE