Autor: |
F. Bourhaleb, Cristiana Peroni, Flavio Marchetto, G. Pittà, F. Martin, S. Meyroneinc, Roberto Cirio, Marco Donetti, M.A. Garella, N. Givehchi, A. La Rosa, S. Giordanengo |
Rok vydání: |
2006 |
Předmět: |
|
Zdroj: |
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 565:833-840 |
ISSN: |
0168-9002 |
DOI: |
10.1016/j.nima.2006.06.024 |
Popis: |
The Dipartimento di Fisica Sperimentale and the Istituto Nazionale di Fisica Nucleare (INFN), Torino, in collaboration with the Institut Curie—Centre de Protontherapie de Orsay (CPO), have developed and built a pixel parallel plate ionization chamber to be used as monitor for the proton therapy beam line at the Institut Curie—CPO (Orsay, France). The sensitive area of the detector is (160×160) mm2, with the anode segmented in 1024 square pixels arranged in a 32×32 matrix; the area of each pixel is (5×5) mm2. The detector has been placed on the beam line just upstream of the last collimator to monitor the beam shape and to measure the stability and reproducibility of the delivery system. In this paper, we present a detailed description of the detector and the results of a set of preliminary tests. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|