Popis: A comparison has been made between a graphite furnace system based on nickel as a matrix stabilizing metal and an automated hydride generation system with a heated quartz cell. The effect of nickel as a matrix modifier was studied in pure selenite solutions as well as in biological matrixes by different charring temperatures. The suppression effect of different acids on the response of the analyte is reported and discussed. The use of an electrically heated quartz tube as an alternative to the argon hydrogen flame method unproved the selenium determination by hydride generation atomic absorption. The effect of hydrochloric acid to secure quantitative formation of selenium (IV) and the interference of copper in the response measurements have been studied. Further a comparison has been made between three different digestion procedures when the hydride generation atomic absorption system was applied. The results of the graphite furnace atomic absorption and the hydride generation atomic absorption were found to be equally accurate, but the graphite furnace technique gave better reproducibility.