A thermally polarized 129 Xe phantom for quality assurance in multi‐center hyperpolarized gas MRI studies
Autor: | Ralph Hashoian, Bastiaan Driehuys, E. Bier, Ziyi Wang, Geoffry M. Schrank, Mu He, John Nouls, Naomi Morales‐Medina, Harvey Svetlik, John P. Mugler |
---|---|
Rok vydání: | 2019 |
Předmět: |
Scanner
Materials science Pixel chemistry.chemical_element Imaging phantom 030218 nuclear medicine & medical imaging 03 medical and health sciences Laser linewidth 0302 clinical medicine Xenon chemistry Flip angle Electromagnetic coil Radiology Nuclear Medicine and imaging 030217 neurology & neurosurgery Excitation Biomedical engineering |
Zdroj: | Magnetic Resonance in Medicine. 82:1961-1968 |
ISSN: | 1522-2594 0740-3194 |
Popis: | PURPOSE Hyperpolarized 129 Xe MR is increasingly being adopted worldwide, but no standards exist for assessing or comparing performance at different 129 Xe imaging centers. Therefore, we sought to develop a thermally polarized xenon phantom assembly, approximating the size of a human torso, along with an associated imaging protocol to enable rapid quality-assurance imaging. METHODS MR-compatible pressure vessels, with an internal volume of 5.85 L, were constructed from pressure-rated, engineering grade PE4710 high-density polyethylene. They were filled with a mixture of 61% natural xenon and 39% oxygen to approximately 11.6 bar and placed in a loader shell filled with a 0.56% saline solution to mimic the human chest. Imaging employed a 2D spoiled gradient-echo sequence using non-slice-selective excitation (TR/TE = 750/6.13 ms, flip angle = 74°, FOV = 40 × 440 mm, matrix = 64 × 32, bandwidth = 30 Hz/pixel, averages = 4), resulting in a 1.6 min acquisition. System characterization and imaging were performed at 8 different MRI centers. RESULTS At 3 Telsa, 129 Xe in the pressure vessels was characterized by T1 = 580.5 ± 8.3 ms, linewidth = 0.21 ppm, and chemical shift = +10.2 ppm. The phantom assembly was used to obtain transmit voltage calibrations and 2D and 3D images across multiple coil and scanner configurations at 8 sites. Across the 5 sites that employed a standard flexible chest coil, the SNR was 12.4 ± 1.8. CONCLUSION The high-density polyethylene pressure vessels filled with thermally polarized xenon and associated loader shell combine to form a phantom assembly that enables spectroscopic and imaging acquisitions that can be used for testing, quality assurance, and performance tracking-capabilities essential for standardizing hyperpolarized 129 Xe MRI within and across institutions. |
Databáze: | OpenAIRE |
Externí odkaz: |