Vergleich eines High-End-Ultraschallgerätes mit einem Spiral-CT-Scanner beim Screening von Lebermetastasen

Autor: B. Wintersperger, Harald Sittek, Harald M. Bonel, M. F. Reiser, M. D. Seemann, Kathrin Herrmann
Rok vydání: 2008
Předmět:
Zdroj: Ultraschall in der Medizin. 19:164-167
ISSN: 1438-8782
0172-4614
DOI: 10.1055/s-2007-1000483
Popis: AIM To assess prospectively the sensitivity of a high-end ultrasound system in comparison to a spiral computed tomographic scanner in screening for liver metastases. METHOD 88 consecutive patients with malignant diseases were examined using gray-scale sonography and contrast enhanced two-phase spiral computed tomography during arterial dominant and portal venous phase. For both methods all focal lesions of the liver were reported. RESULTS 99 focal lesions of the liver were detected by gray-scale sonography and/or two-phase spiral computed tomography. The sensitivity for gray-scale sonography was 93.9% (93 of 99 lesions) and for two-phase spiral computed tomography (71.7% (71 of 99 lesions), respectively. CONCLUSION Screening for liver metastases should be performed by using a diagnostic method with a high sensitivity. Gray-scale sonography with a high-end ultrasound system fulfill' these requirements.