Characterization and use of a gamma radiation TLC scanner for radiochemical purity measurements of radiopharmaceutical kit preparations of99mTc(dmpe)2Cl 2 +

Autor: S. I. Sivakoff, A. P. Carpenter, V. Subramanyam, L. L. Camin, T. J. Mysliwy, E. M. Wolshin, M. A. Matthew, G. H. Schnaper, M. N. Doshi
Rok vydání: 1985
Předmět:
Zdroj: Journal of Radioanalytical and Nuclear Chemistry. 89:435-446
ISSN: 1588-2780
0236-5731
DOI: 10.1007/bf02040607
Popis: A simple, low cost TLC scanner system for the characterization of γ-emitting radiopharmaceuticals is described. The TLC γ scanner has a linear response range of ∼0.074 Mbq up to ≥2.94 Mbq. The accuracy and reproducibility of measurement is better than ±5% (relative). The utility of this TLC scanner system is demonstrated by application to the radioanalytical evaluation of99mTc(dmpe)2Cl 2 + preparations. A comparison with HPLC radioanalytical separations is made.
Databáze: OpenAIRE