A high sensitivity and strong stability miniaturized optically-pumped magnetometer
Autor: | Zinan Wu, Yang Zhang, Mengyang He, Bokang Ren, Huimin Yue, Zhonghua Ou, Xiaojun Zhou, Yong Liu |
---|---|
Rok vydání: | 2022 |
Předmět: | |
Zdroj: | Japanese Journal of Applied Physics. 61:070904 |
ISSN: | 1347-4065 0021-4922 |
DOI: | 10.35848/1347-4065/ac748c |
Popis: | In this paper, we focus on the phase noise in optically-pumped magnetometer systems and investigate its sources, and use dual-phase lock-in technique to solve the problem. The results show that the phase noise is maximum when the system operates under zero-magnetic-field condition, and the main source comes from the environment stray magnetic field. The dual-phase lock-in technique improves the phase noise rejection by a factor of 10 compared to the single-phase lock-in technique. The sensitivity of the system is 22 fT Hz−1/2 and the bandwidth is 158 Hz, which offers a favorable outlook for use in clinical settings. |
Databáze: | OpenAIRE |
Externí odkaz: |