k 0-INAA of biological matrices at IPEN neutron activation analysis laboratory, São Paulo, using the k0_IAEA software

Autor: R. Semmler, D. C. Puerta, Ana Maria Graciano Figueiredo, R. Jaćimović
Rok vydání: 2014
Předmět:
Zdroj: Journal of Radioanalytical and Nuclear Chemistry. 300:637-644
ISSN: 1588-2780
0236-5731
Popis: This study presents the results obtained in the application of the k 0-standardization method at the Neutron Activation Analysis Laboratory at IPEN (LAN-IPEN), for biological sample analysis, by using the k0_IAEA software, provided by the International Atomic Energy Agency (IAEA). The thermal to epithermal flux ratio f and the shape factor α of the epithermal flux distribution of the IEA-R1 nuclear reactor of IPEN were determined for the pneumatic irradiation facility and one selected irradiation position, for short and long irradiations, respectively. To obtain these factors, the “bare triple-monitor” method with 197Au–96Zr–94Zr was used. To evaluate the accuracy of the results, bias (%) and E n-number test were applied to the results obtained in the analysis of the biological reference materials NIST SRM 1547 peach leaves, INCT-MPH-2 mixed polish herbs and NIST SRM 1573a tomato leaves. Bias (%), for most elements, ranged from 0 to 30 %, in relation to certified values. E n-number values showed that, with few exceptions (Na in NIST SRM 1547 and NIST SRM 1573a, and Al, Cr, Sc and Zn in INCT-MPH-2), the results were within a 95 % CI. These results pointed to the possibility of using the k 0-INAA method with the k0_IAEA software for analysis of biological samples at LAN-IPEN.
Databáze: OpenAIRE