New air speed calibration system at NMIJ for air speeds up to 90 m/s

Autor: Yoshiya Terao, Aya Iwai, Yong Moon Choi, Tatsuya Funaki, Noboru Kurihara
Rok vydání: 2019
Předmět:
Zdroj: Flow Measurement and Instrumentation. 66:132-140
ISSN: 0955-5986
Popis: National Metrology Institute of Japan (NMIJ) has established a high air speed standard facility and has been providing a calibration service since 2015. The facility has an air speed range of 40 m/s to 90 m/s with a relative expanded uncertainty (k = 2) of 0.63%. The purpose of this primary standard is mainly to contribute to the improvement of meteorological observation research and to the evaluation of flow field inside a turbo machine and around a high speed vehicle. The reference air speed is derived from the national primary gas flowrate standard of Japan. A conversion device from flowrate to air speed is installed in the test line of the closed-loop calibration facility. The reference air speed at the nozzle exit of the conversion device is obtained by comparing the integral of the air speed profiles and the reference volume flowrate. The total pressure tube used as a transfer standard is then calibrated against the reference air speed at the center of the nozzle exit. The Eiffel-type wind tunnel, which is a working standard for the daily calibration service, is calibrated using this total pressure tube. The present paper describes the calibration system, the traceability chain, and an uncertainty analysis using a validation method.