Postimplantation Changes in Components of Hip Endoprostheses
Autor: | K. K. Karimov, A. R. Trubin, G. N. Filimonov, T. E. Khairov, R. R. Yakupov, A. V. Sisanbaev, I. R. Gafarov, K. H. Sirodzhov |
---|---|
Rok vydání: | 2016 |
Předmět: |
030222 orthopedics
Insert (composites) Materials science 3d analysis 0206 medical engineering Biomedical Engineering Medicine (miscellaneous) 02 engineering and technology Kinematics Tribology 020601 biomedical engineering 03 medical and health sciences Medical Laboratory Technology 0302 clinical medicine Eccentric Implant Joint (geology) Surface deformation Biomedical engineering |
Zdroj: | Biomedical Engineering. 50:58-62 |
ISSN: | 1573-8256 0006-3398 |
Popis: | Surface deformation relief (SDR) of the contact areas of a hip joint endoprosthesis subjected to wear was studied at different structural levels using laser scanning microscopy. Quantitative comparative 3D analysis of the SDR parameters characterizing postimplantation changes in components of the implant was performed. It was found that wear of the kinematic unit was eccentric and heterogeneous. Deep local surface changes were mainly observed at the upper pole of the head and the insert. The bottom segments of the insert remained intact even after long-term use of the tribological pair. The SDR of these segments remained intact. |