An improved ion source for thermal ionization mass spectrometer

Autor: E. Ravisankar, V. Nataraju, T.K. Saha, Varun K. Yadav, R.K. Bhatia, S.C. Gadkari, Arkadip Basak
Rok vydání: 2017
Předmět:
Zdroj: International Journal of Mass Spectrometry. 423:33-38
ISSN: 1387-3806
DOI: 10.1016/j.ijms.2017.10.006
Popis: The ion source of a thermal ionization mass spectrometer (TIMS) has been designed to enhance the source sensitivity by modifying the geometry of extraction lens. Computer simulation studies on different combinations of the electro-static lens components and experimental validations were carried out. This also involved designing a robust triple filament assembly in a home-built TIMS. A good agreement between the simulated and experimental results, led to overall improvements in terms of the consistency of the source performance, a higher through put and an overall sensitivity enhancement by a factor of two.
Databáze: OpenAIRE