Recent Progress in Cs Beam Frequency Standards at the NRLM

Autor: Yasuki Koga, Yasuhiro Nakadan
Rok vydání: 1985
Předmět:
Zdroj: IEEE Transactions on Instrumentation and Measurement. :133-135
ISSN: 1557-9662
0018-9456
DOI: 10.1109/tim.1985.4315286
Popis: Some causes and cures of frequency uncertainty in the cesium beam frequency standard, NRLM-II, are described. Accuracy reevaluation indicates about 2.2 × 10-13, and the standard frequency is about 8 × 10-14 below that of the International Atomic Time (TAI) scale. The frequency stability is estimated as better than 8 × 10-12 ?-1/2 for short-term and 4 × 10-14 for 1-2 day averages.
Databáze: OpenAIRE