New dip-and-read test for determining leukocytes in urine
Autor: | J A Cattell, L R Stover, Paul F Corey, J H Pendergrass, R. W. Trimmer, J P Albarella, R J Tyhach, A. E. Burkhardt, A C Skjold, M P Farinelli |
---|---|
Rok vydání: | 1987 |
Předmět: | |
Zdroj: | Clinical Chemistry. 33:1242-1245 |
ISSN: | 1530-8561 0009-9147 |
DOI: | 10.1093/clinchem/33.7.1242 |
Popis: | A new reagent strip for the determination of leukocytes in urine (LEUKOSTIX; Ames) is described. The test is based on the esterase activity in leukocytes as a marker. Upon contact between the reagent matrix and a urine containing leukocytes, an amino acid ester is hydrolyzed by the esterase to its corresponding alcohol. The free alcohol then couples with a diazonium salt to produce a purple azo dye. The relative concentration of leukocytes in the urine is obtained by visually comparing the strip reaction with a color chart. Performance of the strip was evaluated in a clinical study involving eight different sites and 867 urine specimens. The comparison method was sediment microscopy; specimens containing five cells or more per high-power field were considered to be positive. Sensitivity was 76.3%, specificity 80.8%. Performance was comparable with that of the CHEMSTRIP LN (Boehringer-Mannheim Diagnostics, Inc.) leukocyte test, which we evaluated concurrently. |