Algorithms For Limited-Angle Computed Tomography
Autor: | Douglas A. Ortendahl, Douglas P. Boyd, William H. Rowan, John L. Couch |
---|---|
Rok vydání: | 1982 |
Předmět: |
Artifact (error)
Scanner medicine.diagnostic_test business.industry Computer science Optical engineering ComputingMethodologies_IMAGEPROCESSINGANDCOMPUTERVISION Orthogonal functions Computed tomography Imaging phantom Convolution Limited angle medicine Computer vision Artificial intelligence business Algorithm ComputingMethodologies_COMPUTERGRAPHICS |
Zdroj: | SPIE Proceedings. |
ISSN: | 0277-786X |
DOI: | 10.1117/12.934510 |
Popis: | Limited-angle computed tomography was studied in a project to develop algorithms for a limited-angle scanner. Both the ART algorithm and an orthogonal function algorithm were investigated. The artifact produced by the different methods was very similar. A method is presented for producing model limited-angle artifact in phantom images. Filters are investigated for reducing the artifact, and it is shown that substantial improvement in subjective quality of images can be obtained. This method can be incorporated into limited-angle convolution back-projection algorithms.© (1982) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only. |
Databáze: | OpenAIRE |
Externí odkaz: |