Characterization of the Teledyne CHROMA HgCdTe detector for imaging spectrometers
Autor: | Michael Eastwood, Elliott Liggett, Peter Sullivan, Robert O. Green, Michael Bernas |
---|---|
Rok vydání: | 2017 |
Předmět: |
Physics
Spectrometer business.industry Dynamic range Detector Imaging spectrometer Hyperspectral imaging Context (language use) 02 engineering and technology 021001 nanoscience & nanotechnology 01 natural sciences 010309 optics Imaging spectroscopy Optics 0103 physical sciences 0210 nano-technology business Dark current |
Zdroj: | 2017 IEEE Aerospace Conference. |
Popis: | Imaging spectroscopy places high demands upon detector performance. The dynamic range, linearity, and sensitivity must be maintained at high frame rates, and artifacts must be minimized across large focal plane array (FPA) formats. In this contribution, we discuss the Teledyne CHROMA HgCdTe FPA in the context of imaging spectrometer applications. The FPA was characterized in a laboratory thermal-vacuum chamber that allowed the read noise, dark current, well capacity, linearity, and crosstalk to be measured. Results are presented across a range of operating temperatures and pixel clock rates. Additionally, the outcomes of radiation testing to 100 krad TID are discussed. The results of these characterizations are critical to the designs and performance predictions of future imaging spectrometers utilizing the HgCdTe CHROMA FPA. |
Databáze: | OpenAIRE |