Evaluation of Spatial Resolution of MRI, Optical CT and X-Ray CT Using MTF for Gel Dosimeter
Autor: | Takaoki Takanashi, Hiraku Kawamura |
---|---|
Rok vydání: | 2018 |
Předmět: | |
Zdroj: | IFMBE Proceedings ISBN: 9789811090349 |
DOI: | 10.1007/978-981-10-9035-6_6 |
Popis: | We have proposed a new gelatin phantom for evaluating the spatial resolution of Optical Computed Tomography (OCT), MRI and X-ray CT (X-CT) as a gel dosimeter readout device. In addition, MTF evaluation was conducted using the proposed gelatin phantom. Consequently, 0.19 (MRI), 1.07 (X-CT), 0.88 (0-D OCT), and 0.89 (2-D OCT) were obtained as a results of modulation transfer factor at 0.3 spatial frequency (cycle/mm). |